Open original FDA record ↗Recall record
- FDA identifier
- Z-1604-2023
- Company in FDA record
- Hiossen Inc.
- FDA classification date
- May 18, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Mis-labeled length of the dental implant on the Recall Implant's package. Dental implant size D5.0 x L8.5mm, labeled as D5.0 x L10mm
Source facts
- Action
- Hoissen Inc. initial contact verbally through sales rep. on April 7, 2023 and by "URGENT: MEDICAL DEVICE RECALL" Letter via Mail & email on 04/19/2023. Letter states reason for recall, health risk and action to take: 1. Do not use the Recall Implants. 2. Quarantine (separate) the Recall Implants in a different location, away from normal stock room. 3. Immediately fill out the Acknowledgement and Receipt Form* (Fax: 267-795-7031 or Email: recall@hiossen.com) *Response is required regardless if the Recall Implants have already been return or implanted. 4. Returned the Recall Implant(s) as Unusable Returns to Hiossen QA** **Contact your sales representative for assistance. 5. Replacement devices are available for delivery. If you need any further information or support concerning this recall/issue, please reach HIOSSEN: Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time. Tel: 267-795-7031 Email: recall@hiossen.com
- Initiated
- 2023-04-06
- Root cause
- Labeling Change Control
- Distribution
- US Nationwide Distribution
- Recall number
- Z-1604-2023
- Recall status
- Open, Classified
- CDRH record ID
- 200035
- FDA classified
- 2023-05-18
- Recall event number
- 92272