Skip to content

Public FDA record

Z-1605-2023: HydroPICC 4Fr Single Lumen Catheter, Maximal Barrier Kit, Model Number 80001004

Recall record Z-1605-2023 for Access Vascular, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1605-2023
Company in FDA record
Access Vascular, Inc
FDA classification date
May 18, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The catheter packaged with specific lots of the HydroPICC Kits and HydroMID Kits have a shorter expiration date than what is listed on the outer kit package.

Source facts

Action
An "Urgent Medical Device Correction" notification letter was sent to customers on 3/3/23. Our records indicate that your facility has received kits from the affected lots. We are contacting you to provide information on action to take to ensure patient safety. - Product that has already been used: No action required. The catheter and all kit components are currently within their expiration date, so no patient safety risk has been identified. - Product that has not been used: The catheter and all kit components are currently, within their expiration date, so no immediate patient safety risk has been identified. However, please refrain from using these kits as using the product used past its expiration date may contaminate the sterile field and result in an infection. AVI will provide a replacement label with the correct expiration date or replace the affected products. Access Vascular is committed to providing the highest quality products and support and is implementing corrective actions to prevent this issue from recurring in the future. We apologize for any inconvenience this may cause you and appreciate your understanding as we take steps to correct this issue. If you have any questions about this communication, please contact us at 781-538-6594.
Initiated
2023-03-03
Root cause
Labeling mix-ups
Distribution
US distribution to States of: FL, TN, TX.
Recall number
Z-1605-2023
Recall status
Open, Classified
CDRH record ID
199744
FDA classified
2023-05-18
Recall event number
92124

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗