Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
Source facts
Action
On 2/14/24 the firm phoned distributors and informed them of the recall issue and next steps. On 2/16/24 the firm sent a recall communication via email. Customers were asked to arrange for the return of these sets along with any additional Mariner RDX inventory using a firm provided return label.
Customers with questions or concerns can contact the Associate Product Manager, at olivia.disanto@seaspine.com
Initiated
2024-02-15
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of MI, GA, VA, IN, FL.
Recall number
Z-1605-2024
Recall status
Open, Classified
CDRH record ID
206495
FDA classified
2024-04-22
Recall event number
94164
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.