Skip to content

Public FDA record

Z-1606-2008: Dimension Total Prostate Specific Antigen (TPSA) Flex Reagent Cartridge (RF451). Siemens, Newark, DE 19

Recall record Z-1606-2008 for Siemens Healthcare Diagnostics Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1606-2008
Company in FDA record
Siemens Healthcare Diagnostics Inc.
FDA classification date
Jun 28, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

False Results: The product may exhibit falsely elevated results. The affected lots demonstrate sporadic variability in recovery of calibrator and quality control; therefore this issue may not be readily detected upon calibration or during routine processing of Quality Control.

Source facts

Action
The recalling firm issued a letter (Urgent Field Safety Notice) dated 3/25/08 to inform their customers to discontinue use of any remaining inventory, retest using an alternate lot for those samples from post-prostatectomy patients which recovered positive TPSA results less than 1.0 ng/nL. The firm has requested that you call the them for no-charge replacement product at 1-800-441-9250 and/or for technical assistance.
Initiated
2008-03-24
Root cause
Device Design
Terminated
2008-10-15
Distribution
The product was shipped to medical facilities nationwide. The product was also shipped to Germany, Singapore, Canada, Japan and AE
Recall number
Z-1606-2008
Recall status
Terminated
CDRH record ID
69659
FDA classified
2008-06-28
Recall event number
47518

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗