Open original FDA record ↗Recall record
- FDA identifier
- Z-1606-2008
- Company in FDA record
- Siemens Healthcare Diagnostics Inc.
- FDA classification date
- Jun 28, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
False Results: The product may exhibit falsely elevated results. The affected lots demonstrate sporadic variability in recovery of calibrator and quality control; therefore this issue may not be readily detected upon calibration or during routine processing of Quality Control.
Source facts
- Action
- The recalling firm issued a letter (Urgent Field Safety Notice) dated 3/25/08 to inform their customers to discontinue use of any remaining inventory, retest using an alternate lot for those samples from post-prostatectomy patients which recovered positive TPSA results less than 1.0 ng/nL. The firm has requested that you call the them for no-charge replacement product at 1-800-441-9250 and/or for technical assistance.
- Initiated
- 2008-03-24
- Root cause
- Device Design
- Terminated
- 2008-10-15
- Distribution
- The product was shipped to medical facilities nationwide. The product was also shipped to Germany, Singapore, Canada, Japan and AE
- Recall number
- Z-1606-2008
- Recall status
- Terminated
- CDRH record ID
- 69659
- FDA classified
- 2008-06-28
- Recall event number
- 47518