Open original FDA record ↗Recall record
- FDA identifier
- Z-1606-2011
- Company in FDA record
- Fujirebio Diagnostics, Inc.
- FDA classification date
- Mar 8, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
presence of material attached to the bottom of some wells in places visible to the naked eye, which caused an overestimation of the sample.
Source facts
- Action
- Notification dated 7/7/10 was sent to the customer via e-mail and FedEx Priority Overnight. FUJIREBIO Diagnostics, Inc. sent an URGENT PRODUCT CORRECTION NOTIIFICATION letter dated July 7, 2010, to all affected customers via e-mail and FedEx Priority Overnight. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue using MESOMARK kit. Lot 57K04510 and discard any remaining inventory. Complete and return the attached Confirmation of Receipt form upon receipt of this notification, no later than July 30, 2010. Notify your ordering physician or health care provider as appropriate. For questions regarding this recall call 877-861-7246 or 610-240-3800.
- Initiated
- 2010-07-29
- Root cause
- Other
- Terminated
- 2012-07-05
- Distribution
- Nationwide Distribution including UT
- Recall number
- Z-1606-2011
- Recall status
- Terminated
- CDRH record ID
- 93728
- FDA classified
- 2011-03-08
- Recall event number
- 56427