- FDA identifier
- Z-1606-2015
- Company in FDA record
- GE Healthcare
- FDA classification date
- May 12, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The heart rate could be calculated from pacer pulses without indication that pacemaker detection is OFF in some combinations of Bx50 monitors, a PDM, and CIC/CSCS. Then the monitor does not show the Pacer Off message and there may not be an alarm for asystole. Undetected asystole could result in irreversible changes in the patients condition and delayed or missed life sustaining patient treatment
Source facts
- Action
- GE Healthcare sent an Urgent Medical Device Correction letter dated May 1, 2015, to all affected consignees. The letter was addressed to Healthcare Administrator / Risk Manager, Chief of Nursing, and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Initiated
- 2015-05-04
- Root cause
- Software design
- Terminated
- 2015-09-15
- Distribution
- Worldwide Distribution - US including CO, MI, WI and Internationally to NETHERLANDS, FINLAND, and BELGIUM.
- Recall number
- Z-1606-2015
- Recall status
- Terminated
- CDRH record ID
- 136601
- FDA classified
- 2015-05-12
- Recall event number
- 71211