Z-1606-2024: Various Alcon Custom Paks Surgical Procedure Pack containing Oasis 27Ga Chang Hydrodissection Cannulas.

Recall record Z-1606-2024 for Alcon Research, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1606-2024
Company in FDA record
Alcon Research, LLC
FDA classification date
Apr 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to complaints of particulates, residue, occlusions, and bent tips for sterile cannulas.

Source facts

Action
On 03/15/2024, the firm sent an "URGENT: VOLUNTARY MEDICAL DEVICE FIELD CORRECTION" Letter via FedEx overnight to inform customer that, specific lots of Alcon Custom Pak containing Oasis Medical 27G Chang Hydrodissection Cannulas. Alcon has received customer complaints for foreign material, occlusions and bent tips for these cannulas. We are asking that you remove and dispose of the included 27G Chang Hydrodissection Cannula upon opening your Custom Pak for its specific surgical case. Customer are to take the following actions: 1. See Table 1 for list of affected Alcon Custom Pak lots that were distributed to their facility. They may have received Custom Pak lots assembled prior to this Field Correction notification, and as a result, they may be receiving affected inventory in future shipments. Any affected Alcon Custom Pak in the future, will be identified by a sticker and reads, "URGENT FIELD CORRECTION Discard enclosed 27G Chang Hydrodissection Cannula" 2. Review their inventory of Alcon Custom Pak lots against the lots listed in Table 1. For affected lots in their current on-hand inventory, affix the provided stickers directly to the outside of their affected Custom Pak in a location most easily seen. NOTE: Alcon Custom Pak units are sterile and should not be opened prior to surgery. 3. Upon opening of the identified Alcon Custom Pak for surgical use, remove and dispose of the enclosed 27G Chang Hydrodissection Cannula. 4. Use a separate sterile standalone hydrodissection cannula, either from their own inventory or those supplied by Alcon. For impacted products that will ship in the future, Alcon is planning to ship sterile stand-alone substitutions. Call Alcon Customer Service (1-800-862-5266) if help it needed in obtaining replacement sterile standalone hydrodissection cannulas. 5. Keep Table 1 for their records should they receive any affected Alcon Custom Pak lots in the near future. Questions/concerns, contact Alcon Customer Service at 1-800-862-5266.
Initiated
2024-03-15
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - U.S. Nationwide including in the states of AK, AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Australia, Canada, and New Zealand.
Recall number
Z-1606-2024
Recall status
Open, Classified
CDRH record ID
206740
FDA classified
2024-04-22
Recall event number
94253

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗