Z-1607-2018: Mutlidiagnost-Eleva with Flat Detector Eleva with Flat Detector 708037 As a multifunctional universal i

Recall record Z-1607-2018 for Philips Medical Systems Nederlands. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1607-2018
Company in FDA record
Philips Medical Systems Nederlands
FDA classification date
May 3, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Source facts

Action
The firm, BU IGT Systems, sent an "URGENT- Field Safety Notice" letter dated 3/23/2018 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to stop using the system and call your Philips representative if the following situations occurs: -the table height or tilt movement is not working. - you notice a cracking or snapping sound during system operation. - you experience a blockade in the table height or tilt movement during system movements. -the system is having an unexpected collision or the system had an unexpected collision in the past 2 months. - you have used a CPR stand that obstructed the system, or - you notice any (other) unusual system behavior beyond the normal use of the system. The customers should also avoid collisions with the table and base stand of the system by ensuring that no obstacles are placed around the system. In addition, ensure that all staff with access to the affected systems are informed of the content of this Field Safety Notice. The firm stated: -All possibly affected products in the field will be corrected by means of a field change order free of charge. -A firm representative will replace affected material on the upper and lower tilt actuator in the affected systems. -You will be contacted by our local Philips representative to schedule this corrective action. -This action will start effective April 2018. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722- 9377, 8am to 8pm (EST) Monday thru Friday.
Initiated
2018-03-26
Root cause
Nonconforming Material/Component
Terminated
2020-06-12
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Recall number
Z-1607-2018
Recall status
Terminated
CDRH record ID
163437
FDA classified
2018-05-03
Recall event number
79783

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗