Open original FDA record ↗Recall record
- FDA identifier
- Z-1608-2015
- Company in FDA record
- Spacelabs Healthcare Inc
- FDA classification date
- May 13, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact Monitor to fail to turn on. This is a retrospective recall related to a recall initiated in 2012 (Z-0403-2013).
Source facts
- Action
- The firm sent the Urgent Medical Device Correction- Spacelabs Healthcare Inc.-qube Compact Monitor, Model 91390, Power Input PCBA Failure letter, dated 8 May 2015, via priority service mail to US consignees. This letter described the affected system, problem, actions to be taken by users and corrective actions conducted by Spacelabs. For additional information or technical assistance, please contact: Technical Support Spacelabs Healthcare, Inc. 35301 SE Center St. Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support
- Initiated
- 2015-05-08
- Root cause
- Software design
- Terminated
- 2015-11-05
- Distribution
- distributed in North Carolina, Texas, and Wyoming
- Recall number
- Z-1608-2015
- Recall status
- Terminated
- CDRH record ID
- 136920
- FDA classified
- 2015-05-13
- Recall event number
- 71233