Medtronic has identified one lot #0610315 of U-Clip Removal Tools, Catalog UREM1860FW, that was incorrectly assembled. The nose cone attachment, on the body of the removal tool, may not contain sufficient adhesive, which can allow the nose cone to separate from the body of the device.
Source facts
Action
A Medtronic "Urgent Medical Device Field Notification" letter was sent to one distributor and a separate consignee Medtronic letter "Urgent Medical Device Field Notification" was sent to end users on February 25, 2008. The letters explained the situation and requested that the impacted product be returned to Medtronic for disposal. The end user will be requested to complete a Recall certificate acknowledging the receipt of the information.
If you have questions, contact your VNUS Sales Representative or Lauralie Millikan at 763-391-9124.
Initiated
2008-02-25
Root cause
Process control
Terminated
2008-11-10
Distribution
AZ, GA, PA, MI, KY, FL, OH, TX, NY, VA, CA, MO IL.
Recall number
Z-1609-2008
Recall status
Terminated
CDRH record ID
69720
FDA classified
2008-09-16
Recall event number
47538