Skip to content

Public FDA record

Z-1609-2024: t:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology

Recall record Z-1609-2024 for Tandem Diabetes Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1609-2024
Company in FDA record
Tandem Diabetes Care, Inc.
FDA classification date
May 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected. Pump shutdown will cause insulin delivery to suspend, which could lead to an under-delivery of insulin and may result in hyperglycemia, including severe hyperglycemia.

Source facts

Action
On 10/09/2024, Tandem send via email an "URGENT MEDICAL DEVICE CORRECTION UPDATE" informing customers of additional actions required (uninstall and reinstall mobile app version 2.8.2) to address the faster than expected battery depletion issue. Customers are instructed to: 1.Touch and hold the app icon. 2.Tap Remove App, then tap Delete App to confirm. 3.Go to Apple App Store. 4.Search t:connect mobile to locate the t:connect mobile app. 5.Tap Download, or the cloud icon with an arrow. 6.Login with your t:connect account credentials and re-pair with your Tandem t:slim X2 pump. For complete re-pair instructions, please visit our website here. Between the dates of August 9-12, 2024, Tandem issued a Press Release and sent an "URGENT MEDICAL DEVICE CORRECTION UPDATE" Letter due to becoming aware that users may still be experiencing pump battery depletion after updating the mobile app to version 2.7.1. On August 19, 2024, Tandem issued a press release announcing the future release of the next version of the t:connect mobile app version 2.8.2. On August 20, 2024, Tandem sent a follow-up email and in-app push notification to all affected users who Tandem has a valid email announcing the release of T:connect mobile app version 2.8.2. - The customer email communication will include a URL and QR code to provide a method for users to acknowledge receipt. Between the dates of August 21, 2024 - September 4, 2024, Tandem will send an "Urgent Medical Device Correction Update" to all affected app users via First-Class Postal mail. This notification will outline the release of T:connect mobile app version 2.8.2.
Initiated
2024-03-05
Root cause
Software Design Change
Distribution
US nationwide distribution including Puerto Rico.
Recall number
Z-1609-2024
Recall status
Open, Classified
CDRH record ID
206914
FDA classified
2024-05-06
Recall event number
94312

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗