Z-1609-2026: MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in det

Recall record Z-1609-2026 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1609-2026
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Mar 20, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the likely presence of contamination in well(s).

Source facts

Action
On 07/24/2025, the firm sent via mail/email an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Beckman Coulter has received reports that when testing clinical isolates there was no growth observed in Ceftazidime (Caz) 1, 8 and 16 mg/mL wells; however, growth was observed in the 4 mg/mL well with multiple panels from this lot. Upon further investigation, it was confirmed the growth was Bacillus spp. and not the clinical isolates being tested. Customers are instructed to: -Consider reviewing previous Caz antimicrobial test results from this lot for any spot growth detection (e.g., isolates with no growth in in Caz 1, 8 and 16 mg/mL wells and growth in the 4 mg/mL well or if there is an unusual increase in Caz MIC values). -Discontinue use of MicroScan Neg MIC 56 Lot 2026-01-03 For any questions regarding this notice, contact Customer Technical Support: -From our website: http://www.beckmancoulter.com -By phone: call 800 677-7226 option 1 in the United States. If replacement product is needed: -Complete the attached Replacement Order Form and email to askbeckman@beckman.com or fax to (866) 294-7850 OR -Call Client Services at (800) 526-3821
Initiated
2025-07-24
Root cause
Material/Component Contamination
Distribution
U.S. Nationwide distribution in the states of CA, CO, IN, MA, MD, MN, MO, NC, NJ, OK, PA, and TX.
Recall number
Z-1609-2026
Recall status
Open, Classified
CDRH record ID
219076
FDA classified
2026-03-20
Recall event number
98555

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗