Open original FDA record ↗Recall record
- FDA identifier
- Z-1610-2008
- Company in FDA record
- GE Healthcare
- FDA classification date
- Sep 19, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
GE Healthcare was notified that 2 recently installed Precision 500D systems were not in compliance with portions of the Performance Standard for Diagnostic X-ray Equipment. X-ray production was possible from the fluoroscopic X-ray tube when the Primary Protective Barrier was not in position to intercept the X-ray beam. Users are recommended to follow good clinical standards to ensure the Image I
Source facts
- Action
- A Product Safety Notification letter, dated 01/09/2008 was sent to affected customers. This letter described the safety issue, affected product and safety instructions. A GE Field Engineer will schedule a field visit to affected customers to inspect and replace the park latch switch as needed.
- Initiated
- 2008-01-02
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-12-10
- Distribution
- Nationwide, Washington D.C. and Puerto Rico and OUS to include: ARGENTINA, AUSTRALIA, BRAZIL, CANADA, CHILE, DOMINICAN REPUBLIC, EGYPT, GERMANY, HONDURAS, INDIA, INDONESIA, IRELAND, JAMAICA, KOREA, KUWAIT, MALAYSIA, MEXICO, NETHERLANDS, PANAMA, RUSSIAN FEDERATION, SAUDIA ARABIA, SYRIAN ARAB REPUBLIC, TAIWAN, THAILAND, UNITED ARAB EMIRATES, and UNITED KINGDOM,
- Recall number
- Z-1610-2008
- Recall status
- Terminated
- CDRH record ID
- 73494
- FDA classified
- 2008-09-19
- Recall event number
- 49467