Skip to content

Public FDA record

Z-1610-2023: 3M Attest Steam Chemical Integrators, REF 1243B

Recall record Z-1610-2023 for 3M Company - Health Care Business. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1610-2023
Company in FDA record
3M Company - Health Care Business
FDA classification date
May 18, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

3M Attest Steam Chemical Integrators with dates of manufacture between November 6, 2022 and January 17, 2023 have an increased potential to leak ink during a sterilization cycle.

Source facts

Action
3M issued an URGENT: Voluntary Product Recall to its consignees beginning 04/06/2023 via email or USPS certified mail. The notice explained the issue, hazard, and requested that the use of the affected product cease, identify any in stock product and dispose of it. The firm will replace the product. To request product replacement for affected lots disposed, contact 3M Health Care Customer Helpline at 1-800-228-3957, option #7 or by email to 3MFieldSafetyAction@montage72.com. Please note, replacement product can only be provided following 3M receipt of the completed form. If you have questions, please the 3M Health Care Customer Helpline at 1-800-228-3957, option #7.
Initiated
2023-04-06
Root cause
Process control
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Australia, Bolivia, Brazil, Canada, Colombia, Czech Republic, Ecuador, Estonia, Finland, France, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, New Zealand, Poland, Saudi Arabia, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, and Vietnam.
Recall number
Z-1610-2023
Recall status
Open, Classified
CDRH record ID
199455
FDA classified
2023-05-18
Recall event number
92041

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗