Z-1610-2026: Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.

Recall record Z-1610-2026 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1610-2026
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Mar 20, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.

Source facts

Action
Olympus notified consignees on about 02/12/2026 via URGENT: MEDICAL DEVICE CORRECTION letter. Consignees were instructed to immediately cease reprocessing the MAJ-1443 and MAJ-1444 valves in an OER-Pro or OER-Elite and ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification and the full extent of the MAJ-1443 and MAJ-1444 reprocessing steps as detailed in the provided Instructions for Use (IFU). The updated IFU can also be accessed electronically through OlympusConnect.com. Additionally, consignees were instructed to replace any existing copies of the MAJ-1443/MAJ-1444 IFU with the updated IFU provided in this letter, identify and notify all customers if affected units were further distributed and complete and confirm receipt of the notification via the Olympus web portal.
Initiated
2026-02-12
Root cause
Reprocessing Controls
Distribution
US Nationwide distribution.
Recall number
Z-1610-2026
Recall status
Open, Classified
CDRH record ID
218479
FDA classified
2026-03-20
Recall event number
98371

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗