The Mantis Rod Inserter Shaft was found to have deformation of the stainless stell split sphere when subjected to high forces, prohibiting the user from being able to grip the rod implantation and the split sphere may separate and fall into the patient.
Source facts
Action
Stryker Spine issued an "Urgent Product Recall" notification dated April 29, 2009 via Federal Express to all Consignees. Users were instructed to quarantine all affected product and fax a copy of the Product Accountability Form to 201-760-8406. A Stryker sales representative will retrieve the quarantined product.
For further information, contact Stryker Spine at 201-760-8206.
Initiated
2009-04-30
Root cause
Device Design
Terminated
2013-01-23
Distribution
Nationwide Distribution.
Recall number
Z-1612-2009
Recall status
Terminated
CDRH record ID
82079
FDA classified
2009-07-07
Recall event number
52090