Skip to content

Public FDA record

Z-1612-2011: STA-Flex SUBTALAR SCREW, 7 mm x 13 mm, REF 982361, Sterile, Biomet Sports Medicine, 56 East Bell Drive, W

Recall record Z-1612-2011 for Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1612-2011
Company in FDA record
Biomet, Inc.
FDA classification date
Mar 9, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The firm has initiated this recall following internal fatigue testing that shows possible initiation of fatigue cracking in implants under load.

Source facts

Action
The firm sent URGENT MEDICAL DEVICE RECALL NOTICES dated 1/21/2011 to distributors, implanting physicians, and medical facilities. The letter to the distributors and medical facilities states the recalled product should be immediately located and returned to Biomet. In addition, the "FAX Back Response Form" should be completed and returned to the number provided. If product has been further distributed, distributors are to contact hospital personnel via the enclosed "Dear Biomet Customer" notice. The letter to the implanting physicians informs the physician of the recall and lists the patients that were implanted with the device. The physician letter quotes from the surgical technique "the implant should also be removed if any postoperative complications occur". The letter to the implanting physicians does not recommend any action be taken with the devices that are implanted. The notice just makes the surgeons aware of the recall. Distributors and Physicians are to confirm receipt of the letter by contacting 800-348-9500, ext 3755 or 3756. Also, questions related to this matter should be directed to 574-371-3755 or 574-371-3756.
Initiated
2011-01-21
Root cause
Pending
Terminated
2011-11-10
Distribution
Nationwide Distribution.
Recall number
Z-1612-2011
Recall status
Terminated
CDRH record ID
97596
FDA classified
2011-03-09
Recall event number
57831

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗