Under certain circumstances the patient
Demographics in a report exported into the EMR may not match the demographics shown in the corresponding Synapse CV clinical report.
Source facts
Action
Fujifilm Medical on 2/4/14 , contacted users and advised to discontinue using HL7 reports pending FMSU determination of root cause. Firm issued follow-up letters on 2/24/14 and 7/24/14 advising users of a mandatory upgrade to version 5.1.1 I base, in-place upgrade, to correct the problems.
Initiated
2015-02-04
Root cause
Software in the Use Environment
Terminated
2018-09-20
Distribution
Nationwide Distribution including PR and the states of IL, IN, FL, GA, MD, MI, MO, MS, NC, NY, OR, and PA, PR and the country of Canada.
Recall number
Z-1612-2015
Recall status
Terminated
CDRH record ID
135411
FDA classified
2015-05-14
Recall event number
70890