MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component.
Source facts
Action
Philips Healthcare sent an Urgent Medical Device Correction letter dated April 2016, on April 14, 2016, to all affected customers. The letter informed customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue.
The correction will consist of providing a replacement MRx unit, free of charge, for all units affected by this issue. When the replacement unit arrives, please use the return shipping label and packaging provided with the replacement MRx unit to return the affected unit to Philips.
Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: You can continue to use your MRx prior to receiving a replacement unit. But, if you observe a red X in the Ready for Use (RFU) indicator, you should follow the troubleshooting instructions in the MRx IFU.
Customers with questions were instructed to contact their local Philips representative or call 1-800-722-9377.
Initiated
2016-04-14
Root cause
Employee error
Terminated
2019-03-27
Distribution
Worldwide Distribution - US including OH, CA, NC and Internationally to China, Iraq, Japan, and Korea.
Recall number
Z-1612-2016
Recall status
Terminated
CDRH record ID
145058
FDA record created
2016-06-02
Recall event number
73825