Open original FDA record ↗Recall record
- FDA identifier
- Z-1613-2018
- Company in FDA record
- Roche Diagnostics Corporation
- FDA classification date
- May 3, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Source facts
- Action
- ROCHE sent an Urgent Medical Device Correction Letter dated February 22, 2018, via UPS to affected customers. Customers were instructed to complete the following: " 1) Use the Return to Setup and Exit buttons to ensure execution of scheduled AutoQC measurements as described in this Urgent Medical Device Correction (UMDC); 2) Check to see if the next scheduled AutoQC measurement is performed to ensure the AutoQC scheduler is working as expected; 3) If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site; 4) Complete the attached fax form (TP-00146) and fax it to 1-844-491-7776 or scan and email to roche7570@stericycle.com; 5) File this UMDC for future reference. Customers with questions should contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272.
- Initiated
- 2018-02-22
- Root cause
- Software design
- Terminated
- 2019-11-15
- Distribution
- USA ( nationwide)
- Recall number
- Z-1613-2018
- Recall status
- Terminated
- CDRH record ID
- 163024
- FDA classified
- 2018-05-03
- Recall event number
- 79676