Open original FDA record ↗Recall record
- FDA identifier
- Z-1615-2013
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA classification date
- Jun 27, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
While performing interval testing Ortho Clinical Diagnostics (OCD) determined that an excess amount of surfactant was unintentionally added to VITROS Chemistry Products CRP Slides during the manufacturing process.
Source facts
- Action
- Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated May 9, 2013 via FedEx or US Priority Mail to all US customers. Foreign Affiliates were notified by e-mail on May 9, 2013 informing them of the issue. Distributors were instructed to notify their customers of the recall. Customers were instructed to discontinue use, discard the affected product, comp[lete and return the attached Confirmation of Recipt Form, and post this notification by each VITROS System. For questions call Customer Service Center at 1-800-828-6316.
- Initiated
- 2013-05-09
- Root cause
- Employee error
- Terminated
- 2015-06-12
- Distribution
- Worldwide Disribution - USA Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain and Venezuela.
- Recall number
- Z-1615-2013
- Recall status
- Terminated
- CDRH record ID
- 118365
- FDA classified
- 2013-06-27
- Recall event number
- 65247