This lot of the ExploR Modular Radial Stem may be missing the wedge ramp threads. If the wedge ramp threads are missing from the stem, the likelihood of the set screw loosening from the radial stem intraoperatively is increased. If the set screw loosens, a revision surgery may be necessary.
Source facts
Action
Biomet sent an URGENT MEDICAL DEVICE SAFETY ALERT dated April 22, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Initiated
2014-04-22
Root cause
Process control
Terminated
2015-01-13
Distribution
US Nationwide Distribution in the states of NY, MD, NC, SC, MI, OR, and NV.
Recall number
Z-1615-2014
Recall status
Terminated
CDRH record ID
127000
FDA classified
2014-05-13
Recall event number
68144
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.