Open original FDA record ↗Recall record
- FDA identifier
- Z-1615-2015
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- May 14, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain Dimension Integrated Chemistry Systems Instrument Power Configurations may have been installed with a non-standard instrument power plug.The use of a non-standard plug configuration as described may lead to a potential electrical hazard. The likelihood of potential clinical impact or injury due to this issue is extremely low.
Source facts
- Action
- Siemens sent an Urgent Field Safety Notice, dated April 2015, was issued for distribution to affected customers in Japan to notify them of the issue. A Siemens Service Representative will visit customer sites to inspect the Dimension instrument power plug and wall outlet configuration. If the instrument is configured with a non-standard 15 Amp plug/outlet or a 20 Amp twist lock plug, Siemens will correct the issue identified, ensuring that the system is installed according to the instrument requirements.
- Initiated
- 2015-04-02
- Root cause
- Under Investigation by firm
- Terminated
- 2016-08-17
- Distribution
- Distributed in the country of Japan.
- Recall number
- Z-1615-2015
- Recall status
- Terminated
- CDRH record ID
- 135513
- FDA classified
- 2015-05-14
- Recall event number
- 70913