Skip to content

Public FDA record

Z-1615-2024: LightStrike Blackout Curtains, Part numbers; 610-00008, 610-00009, 610-00019, 610-00020, and 610-00024, u

Recall record Z-1615-2024 for XENEX Disinfection Services Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1615-2024
Company in FDA record
XENEX Disinfection Services Inc.
FDA classification date
Apr 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Blackout curtains are no longer recommended as a protective barrier against UV/Ozone exposure to a bystander standing in the same room, but on the other side of the curtain from where a microbial reduction device UV cycle is taking place. Bystanders in the room during the cycle may experience temporary skin reddening, feeling of sand in eyes, nausea, headache, and/or mucous membrane irritation.

Source facts

Action
On 3/18/24, correction notices were emailed to customers informing them of the following: 1) Blackout Curtains can continue to be used to block nuisance light coming from an unoccupied room/area where a UV cycle is being executed. It is not necessary to return curtains. 2) If your practice is to use this Blackout Curtain as a protective barrier, discontinue the use in occupied spaces. 3) If you may have further distributed this product, please identify your customers, and notify them at once of this product field correction or provide us with their contact information so the firm can reach out to them. Complete and return the response form. If you have any questions, call the firm at (830) 832-4299 between the hours of 8:00AM and 5:00PM CST Monday through Friday.
Initiated
2024-03-18
Root cause
Device Design
Distribution
US Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, HI, ID, IL, KY, LA, MI, MN, MO, NC, ND, NV, OH, OK, OR, TN, UT, WA, WV, WY.
Recall number
Z-1615-2024
Recall status
Open, Classified
CDRH record ID
206749
FDA classified
2024-04-23
Recall event number
94258

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗