When using the Advanced SieClear spatial compounding feature a synchronization error may occur which can cause a mismatch between the image and the depth scale.
Source facts
Action
A field corrective action was initiated on 04/25/2012 under Siemens update program US019/12/S in which a "Customer Safety Advisory Notice" was sent to all customers notifying them of the problem and the recommended actions to avoid potential risk. A follow-up field corrective action will be initiated in which the affected software will be updated in the field.
Initiated
2012-04-25
Root cause
Software design
Terminated
2012-10-19
Distribution
Worldwide Distribution -- USA, including the states of TX, NJ, GA and the countries of Germany, United Kingdom, Republic of Korea, Australia, Japan, Poland, Israel, Singapore, France and Saudi Arabia.
Recall number
Z-1616-2012
Recall status
Terminated
CDRH record ID
109121
FDA classified
2012-05-16
Recall event number
61803