Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing Ceftriaxone therapy. The interference is not described in the product labeling.
Source facts
Action
On 4/14/14, Roche Diagnostics Operations issued URGENT MEDICAL DEVICE CORRECTION notification to consignees which included description of the affected product and issue, clinical significance, and actions required.
Consignees are directed to call the ACCU-CHEK Customer Care Service Center at 1-800-858-8072 for any questions about the information contained in the notification.
Initiated
2014-04-14
Root cause
Device Design
Terminated
2014-09-18
Distribution
Nationwide Distribution.
Recall number
Z-1616-2014
Recall status
Terminated
CDRH record ID
126667
FDA classified
2014-05-13
Recall event number
67961