Z-1616-2020: Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 0

Recall record Z-1616-2020 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1616-2020
Company in FDA record
Philips North America, LLC
FDA record creation date
Mar 31, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00

Source facts

Action
Philips issued to U.S. customers letter on 2/20/20 via priority mail. Philips Markets Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall ,health risk and action to e taken: Customer/User section of the Customer Information Letter: Included with this Customer Information letter is an Instruction for Use addendum (Part Number 453564720941; Revision B) which provides correct instructions for accessing the O2 sensor calibration menu, as well as other updated product information. Please ensure that the information contained in the addendum is communicated to all users and servicers and is retained with the MR400 Instructions for Use. Upon fulfilling these actions, please complete and sign the attached Reply Form and return it to Philips as directed. Information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377. The firm identified additional accounts and initiated a mailing of the letter on 3/18/20.
Initiated
2020-02-20
Root cause
Labeling design
Terminated
2021-09-21
Distribution
Worldwide distribution - US Nationwide distribution and countries of Utd Arab Emir, Albania, Austria, Australia, Azerbaijan, Bulgaria, Brazil, White Russia, Canada, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Hungary, Ireland, Israel, India, Iceland, Italy, Jamaica, Japan, Kyrgyzstan, Malaysia, Netherlands, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Uzbekistan, Viet Nam.
Recall number
Z-1616-2020
Recall status
Terminated
CDRH record ID
180059
FDA record created
2020-03-31
Recall event number
85090

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗