- FDA identifier
- Z-1616-2020
- Company in FDA record
- Philips North America, LLC
- FDA record creation date
- Mar 31, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00
Source facts
- Action
- Philips issued to U.S. customers letter on 2/20/20 via priority mail. Philips Markets Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall ,health risk and action to e taken: Customer/User section of the Customer Information Letter: Included with this Customer Information letter is an Instruction for Use addendum (Part Number 453564720941; Revision B) which provides correct instructions for accessing the O2 sensor calibration menu, as well as other updated product information. Please ensure that the information contained in the addendum is communicated to all users and servicers and is retained with the MR400 Instructions for Use. Upon fulfilling these actions, please complete and sign the attached Reply Form and return it to Philips as directed. Information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377. The firm identified additional accounts and initiated a mailing of the letter on 3/18/20.
- Initiated
- 2020-02-20
- Root cause
- Labeling design
- Terminated
- 2021-09-21
- Distribution
- Worldwide distribution - US Nationwide distribution and countries of Utd Arab Emir, Albania, Austria, Australia, Azerbaijan, Bulgaria, Brazil, White Russia, Canada, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Hungary, Ireland, Israel, India, Iceland, Italy, Jamaica, Japan, Kyrgyzstan, Malaysia, Netherlands, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Uzbekistan, Viet Nam.
- Recall number
- Z-1616-2020
- Recall status
- Terminated
- CDRH record ID
- 180059
- FDA record created
- 2020-03-31
- Recall event number
- 85090