Z-1616-2023: To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal an

Recall record Z-1616-2023 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1616-2023
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
May 19, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction.

Source facts

Action
Olympus issued Urgent Medical Medical Device Recall letter on 3/31/23 to Endoscopy Department, Pulmonology Department; Risk Management. . Letter states reason for recall, health risk and action to take: 1. Immediately locate any affected products listed in this communication and quarantine these products. 2. Contact our Olympus Customer Service Team at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. " Go to https://olympusamerica.com/recall and enter the recall number 0422 4. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints, including patient injuries and adverse events, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, to report complaints
Initiated
2023-03-31
Root cause
Employee error
Distribution
Natiowide Foreign: Canada, Germany
Recall number
Z-1616-2023
Recall status
Open, Classified
CDRH record ID
199755
FDA classified
2023-05-19
Recall event number
92129

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗