Z-1616-2024: Biomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an establish

Recall record Z-1616-2024 for EBI, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1616-2024
Company in FDA record
EBI, LLC
FDA classification date
Apr 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment

Source facts

Action
ZimVie issued Urgent Medical Device Correction letter to Patients, Physicians, Sales Representatives on 3/22/24 . Letter states reason for recall, health risk and action to take: 1. Review this notification for awareness of the contents. 2. Before each use, inspect your device s lead wires (black cable) for damage. If found to be damaged, contact Customer Service at 1-800-526-2579 x 6000 for a complimentary replacement. 3. Complete the Acknowledgement and Receipt Form either by scanning the QR code on the postcard included with this letter and completing the online form or mailing back the completed postcard. To complete your Acknowledgement, you will need to include your Unique Patient Reference Number included on the top of the first page of this letter. This acknowledgement should be completed even if you do not have affected product. 4. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-526-2579 x 6000. Customer Service Representatives are available from 8:30 am to 5:00 pm (EST), Monday through Friday. At other times, please leave a clear message for a return call by the next business day. Alternatively, your questions may be emailed to ebi.csorders@zimvie.com.
Initiated
2024-03-22
Root cause
Device Design
Distribution
US Nationwide distribution.
Recall number
Z-1616-2024
Recall status
Open, Classified
CDRH record ID
206845
FDA classified
2024-04-23
Recall event number
94291

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗