Z-1617-2020: TEG Manager software versions used in conjunction with TEG 5000 are 1.1.0, 3.0.0, 4.0.0, 4.1.0 and 4.1.1.

Recall record Z-1617-2020 for Haemonetics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1617-2020
Company in FDA record
Haemonetics Corporation
FDA record creation date
Mar 31, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager test result screen are rounded to the nearest whole number and lead to TEG Manager displaying an out of range alert when the test result is actually in range, or vice versa.

Source facts

Action
Haemonetics sent US customers using a TEG 5000 device and TEG Manager software a customer letter via email on February 19, 2020. This customer letter notified end users that Haemonetics is conducting a voluntary correction of all TEG Manager software used in conjunction with the TEG 5000 device. Letter states reason for recall, health risk and action to take: Haemonetics is developing a patch to correct the issue with the TEG Manager software. The patch is expected to be available in March of 2020. Haemonetics will contact customers when the TEG Manager software patch is available for download and we will follow up with each user to ensure that the patch has been received and implemented. In the meantime, customers using TEG Manager with the TEG 5000 device should continue to follow the instructions in the TEG 5000 user manual, which state that, Before making any decisions, the attendant physician must consider all clinical information about the patients condition, the procedure, and previous therapy. Users should rely on the actual value of each test result which is correctly displayed on the TEG 5000 and TEG Manager and should not make clinical decisions based solely on the out of range alerts displayed by TEG Manager software. Recipients of this notice should complete the attached acknowledgement form in its entirety.
Initiated
2020-02-19
Root cause
Software design
Terminated
2021-06-21
Distribution
US Nationwide distribution.
Recall number
Z-1617-2020
Recall status
Terminated
CDRH record ID
180130
FDA record created
2020-03-31
Recall event number
85120

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗