Open original FDA record ↗Recall record
- FDA identifier
- Z-1618-2020
- Company in FDA record
- Karl Storz Endoscopy
- FDA record creation date
- Mar 31, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has become aware that there is a potential that the glass cover at the distal end of the camera/led lighting on the video laryngoscope could become loose during reprocessing or use. If this occurs during usage, it may result in a delay in procedure and/or require removal from body.
Source facts
- Action
- On 12/6/2019 a "Urgent Medical device Correction" letter was sent to affected customers via mail. In addition, to providing information on the recall, the firm asked customers to take the following actions: 1. Please forward this Safety Information to the appropriate departments in your organization. 2. Please train your employees on the new Instructions for Use, especially chapter 5.1. 3. Please complete the attached acknowledgement form and return it to us. 4. If you have any questions about this communication, or needs additional copies of this Instruction for Use, please contact our Technical Support at 800-421-0837, Ext. 5350, Option 5.
- Initiated
- 2019-12-06
- Root cause
- Labeling Change Control
- Terminated
- 2022-07-01
- Distribution
- Worldwide distribution - US Nationwide and country of Canada.
- Recall number
- Z-1618-2020
- Recall status
- Terminated
- CDRH record ID
- 180058
- FDA record created
- 2020-03-31
- Recall event number
- 85089