Skip to content

Public FDA record

Z-1618-2020: Karl Storz SE & Co. KG, C-MAC Video Laryngoscope MIL # 1, for the following REF numbers: 1. 8403xx serie

Recall record Z-1618-2020 for Karl Storz Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1618-2020
Company in FDA record
Karl Storz Endoscopy
FDA record creation date
Mar 31, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has become aware that there is a potential that the glass cover at the distal end of the camera/led lighting on the video laryngoscope could become loose during reprocessing or use. If this occurs during usage, it may result in a delay in procedure and/or require removal from body.

Source facts

Action
On 12/6/2019 a "Urgent Medical device Correction" letter was sent to affected customers via mail. In addition, to providing information on the recall, the firm asked customers to take the following actions: 1. Please forward this Safety Information to the appropriate departments in your organization. 2. Please train your employees on the new Instructions for Use, especially chapter 5.1. 3. Please complete the attached acknowledgement form and return it to us. 4. If you have any questions about this communication, or needs additional copies of this Instruction for Use, please contact our Technical Support at 800-421-0837, Ext. 5350, Option 5.
Initiated
2019-12-06
Root cause
Labeling Change Control
Terminated
2022-07-01
Distribution
Worldwide distribution - US Nationwide and country of Canada.
Recall number
Z-1618-2020
Recall status
Terminated
CDRH record ID
180058
FDA record created
2020-03-31
Recall event number
85089

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗