Open original FDA record ↗Recall record
- FDA identifier
- Z-1619-2010
- Company in FDA record
- I2s Micro Implantable Systems Llc
- FDA classification date
- May 20, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Component of cranial array may not allow sufficient tightening to maintain inserter position.
Source facts
- Action
- The firm, Blackrock Microsystems, sent a "RECALL" letter dated March 26, 2010, to all customers. The letter describes the product, problem and actions to be taken by the customers. The customers were instructed to discontinue use of all affected devices and promptly return any inventory, perform a physical account of your inventory for the recalled product and email a list of all serial numbers to Andrew Gotshalk at agotshalk@blackrockmicro.com, and to please email Andrew Gotshalk even if you do not have any inserter wand holders or have destroyed or removed from use the inserter wand holders. Blackrock will rework or replace all Inserter Wand Holders that is returned under this recall. If you have any questions regarding this recall please call 801-582-5533 x 275.
- Initiated
- 2010-03-26
- Root cause
- Device Design
- Terminated
- 2010-05-25
- Distribution
- Nationwide distribution: CA, MA, NY, RI, and UT.
- Recall number
- Z-1619-2010
- Recall status
- Terminated
- CDRH record ID
- 90518
- FDA classified
- 2010-05-20
- Recall event number
- 55304