Skip to content

Public FDA record

Z-1619-2020: C-Reactive Protein Kit for use on the SPAplus, REF NK044.S.A, Rx only, UDI (01)05051700018456,

Recall record Z-1619-2020 for The Binding Site Group, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1619-2020
Company in FDA record
The Binding Site Group, Ltd.
FDA classification date
Mar 31, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A reagent in C reactive protein kits is not performing to specification and samples with known concentrations less-than-or-equal to 10 mg/L may be affected

Source facts

Action
On 4/25/2019, a Customer Notice was issued by the firm's UK office to firm's US office. The US office was instructed to provide this information to the end customer within 24 hours of receiving the notification along with a form to fill out and return. The Customer Notice issued to the affected customer advises immediate quarantine and disposal of affected kits following local regulations. Customers were asked to pass on the customer notice and to all those who need to be aware within their organization or to any organization where potentially affected devices have been transferred. The customer notice requests the customer to complete, sign and return the e-back form. This confirms receipt of the Customer Notice, the number of kits disposed of and that the advice has been followed, along with a witness signature. Firm requests that e-back form be returned to Technical.support@bindingsite.com or to a local firm representative within one week. Should you require additional information, contact your local representative or Technical Support Group at UK: +44(0) 1214569696.
Initiated
2019-04-25
Root cause
Nonconforming Material/Component
Terminated
2021-04-27
Distribution
U.S.: FL
Recall number
Z-1619-2020
Recall status
Terminated
CDRH record ID
179618
FDA classified
2020-03-31
Recall event number
84890

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗