Open original FDA record ↗Recall record
- FDA identifier
- Z-1620-2010
- Company in FDA record
- Bio-Rad Laboratories
- FDA classification date
- May 20, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The package insert provided with this kit contains a typographical error, giving incorrect directions to prepare a Working Conjugate Solution. If these directions are followed, it would result in a 1:6 dilution and invalidate the assay.
Source facts
- Action
- All domestic customers were notified via an Urgent Product Correction customer letter sent by DHL or FedEx on 8/11/08. Enclosed with the letter was a corrected package insert. The Canadian subsidiary was notified via e-mail on 8/11/08. The letter advised customers of the correct wording, and advised that a new corrected package insert was enclosed and to make note of the correction in the laboratory procedures. Questions should be directed to Bio-Rad Laboratories Technical Support at 1-800-224-6723.
- Initiated
- 2008-08-11
- Root cause
- Packaging process control
- Terminated
- 2010-05-20
- Distribution
- Worldwide Distribution -- USA and Canada.
- Recall number
- Z-1620-2010
- Recall status
- Terminated
- CDRH record ID
- 72844
- FDA classified
- 2010-05-20
- Recall event number
- 49126