Open original FDA record ↗Recall record
- FDA identifier
- Z-1620-2011
- Company in FDA record
- BD Biosciences, Systems & Reagents
- FDA record creation date
- Jun 25, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
To prevent a needle stick while replacing the probe, the firm has initiated a Field Correction, and issued an improved Users Guide procedure (User Bulletin, Part Number 645050).
Source facts
- Action
- The firm, BD Biosciences, sent an information letter dated December 2008 with an enclosed New Probe Replacement Service/User Bulletin, to the customers. The letter described the product, problem and action to be taken. The customers were instructed to review the attached User Bulletin (Part Number 645050); ensure that their laboratory follows the updated procedure for removing and replacing probe; place the User Bulletin in their BD FACS SPA or SPA II User's Guide; always follow the instructions in User Bulletin to properly remove and replace probe, and complete and return the Effectivity Form via fax to 408-577-1409. Should you have any comments or questions, please contact BD Biosciences at 1-877-232-8995 in the US. Outside the US, contact your regional BD office or field service engineer for assistance.
- Initiated
- 2009-01-03
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-03-10
- Distribution
- Worldwide distribution: USA and countries including: Hong Kong, Australia, Belgium, Canada, El Salvador, Hong Kong, Japan, and Mexico.
- Recall number
- Z-1620-2011
- Recall status
- Terminated
- CDRH record ID
- 79200
- FDA record created
- 2015-06-25
- Recall event number
- 51208