Open original FDA record ↗Recall record
- FDA identifier
- Z-1620-2012
- Company in FDA record
- Diamedix Corporation
- FDA classification date
- May 21, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
On 06/08/2011 Diamedix Corporation initiated a recall of Is-ENA-6 Screen Test Kit, Catalog # 720-320 Master Lot S0000, Exp. Date May 31, 2012. The labeling on the kit box has an expiration date of 2012-05-31 while the components have an expiration date of 2012-04-30 for an incorrect expiration date.
Source facts
- Action
- Diamedix Corporation sent a Customer notification letter dated June 8, 2011, to all affected customers via Certified Mail. Customers were instructed to search their inventory and to return affected product to Diamedix. Diamedix apologizes for any inconvenience this has caused our valued customers. Please contact the Diamedix Technical Service Department at 1-800-327-4565 or 305-324-2300 to arrange for the return and replacement of any kits you might still have of the lot in question. Please rest assured that Diamedix values you as a customer and is in the process of taking the proper corrective and preventive actions.
- Initiated
- 2011-06-08
- Root cause
- Error in labeling
- Terminated
- 2013-11-15
- Distribution
- Worldwide Distribution -- (USA) nationwide including the states of FL, NY, NC, MS, KY, MA and Puerto Rico and the countries of Italy, Chili and Venezuela.
- Recall number
- Z-1620-2012
- Recall status
- Terminated
- CDRH record ID
- 108435
- FDA classified
- 2012-05-21
- Recall event number
- 61568