- FDA identifier
- Z-1620-2020
- Company in FDA record
- Stryker Neurovascular
- FDA record creation date
- Mar 31, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has become aware of a potential problem that their Catheters may break/fracture and/or kink/bend if improperly removed from the packaging/dispenser hoop or during catheter use.
Source facts
- Action
- On 04/11/19, the firm sent an "Urgent: Medical Device Correction" to customers via FedEx with tracking and email to inform them the firm has become aware of a potential problem with its AXS Vecta 71 and 74 Aspiration Catheters may fracture if improperly removed from the packaging. If a user does not pull the catheter straight out of the packaging clip, this may introduce a bend onto the proximal section of the catheter. By applying excessive force, the used can cause a break and/or fracture the catheter shaft. Removing the AXS Vecta Aspiration Catheter correctly - not bending the catheter shaft when removing it from the packaging clip, will help to avoid these issues. This recall affects all lots associated with the AXS Vecta 71 Aspiration Catheter (115cm, 125cm, 132cm) and AXS Vecta 74 Aspiration Catheter (115cm, 125cm, 132cm). The firm is instructing customer to: 1. Not to return this product but to ensure the instructions provided is followed to remove the catheter from its packaging. 2. Circulate this Notice internally to all interested/affected parties. 3. Maintain awareness of this communication internally. 4. Inform the firm if any of the subject devices have been distributed to other organizations. 5. Inform the firm of an adverse events concerning the use of the subject devices. 6. Complete and email the customer response form to acknowledge receipt of the communication to NVCustomerCare@Stryker.com.
- Initiated
- 2019-04-10
- Root cause
- Component design/selection
- Terminated
- 2021-02-25
- Distribution
- US Nationwide distribution in the states of PA, FL, CA, IL, WI, AZ, AR, FL, TN, MI, MA, CA, NC, WV, DE, OH, NJ, MO, NY, SC, TX, AL, VA, UT, MD, GA, LA, CO, WA, KY, OK, RI, SD, MS, NE,IA, KS, DC. No US Govt. No Foreign countries.
- Recall number
- Z-1620-2020
- Recall status
- Terminated
- CDRH record ID
- 179354
- FDA record created
- 2020-03-31
- Recall event number
- 84780