Z-1620-2020: Stryker Neurovascular, AXS Vecta 71 Aspiration Catheter, Ref. Catalog No. INC-11129-115, Use By 2021-11-2

Recall record Z-1620-2020 for Stryker Neurovascular. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1620-2020
Company in FDA record
Stryker Neurovascular
FDA record creation date
Mar 31, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has become aware of a potential problem that their Catheters may break/fracture and/or kink/bend if improperly removed from the packaging/dispenser hoop or during catheter use.

Source facts

Action
On 04/11/19, the firm sent an "Urgent: Medical Device Correction" to customers via FedEx with tracking and email to inform them the firm has become aware of a potential problem with its AXS Vecta 71 and 74 Aspiration Catheters may fracture if improperly removed from the packaging. If a user does not pull the catheter straight out of the packaging clip, this may introduce a bend onto the proximal section of the catheter. By applying excessive force, the used can cause a break and/or fracture the catheter shaft. Removing the AXS Vecta Aspiration Catheter correctly - not bending the catheter shaft when removing it from the packaging clip, will help to avoid these issues. This recall affects all lots associated with the AXS Vecta 71 Aspiration Catheter (115cm, 125cm, 132cm) and AXS Vecta 74 Aspiration Catheter (115cm, 125cm, 132cm). The firm is instructing customer to: 1. Not to return this product but to ensure the instructions provided is followed to remove the catheter from its packaging. 2. Circulate this Notice internally to all interested/affected parties. 3. Maintain awareness of this communication internally. 4. Inform the firm if any of the subject devices have been distributed to other organizations. 5. Inform the firm of an adverse events concerning the use of the subject devices. 6. Complete and email the customer response form to acknowledge receipt of the communication to NVCustomerCare@Stryker.com.
Initiated
2019-04-10
Root cause
Component design/selection
Terminated
2021-02-25
Distribution
US Nationwide distribution in the states of PA, FL, CA, IL, WI, AZ, AR, FL, TN, MI, MA, CA, NC, WV, DE, OH, NJ, MO, NY, SC, TX, AL, VA, UT, MD, GA, LA, CO, WA, KY, OK, RI, SD, MS, NE,IA, KS, DC. No US Govt. No Foreign countries.
Recall number
Z-1620-2020
Recall status
Terminated
CDRH record ID
179354
FDA record created
2020-03-31
Recall event number
84780

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗