Recall has been initiated due to concerns regarding the sterility of the product. The firm initiated the recall because of use of inadequately sterilized product might result in patient infection,
Source facts
Action
Reimser, Inc. sent an "URGENT PRODUCT RECALL-COLLAGUIDE COLLAGEN DENTAL MEMBRANE" letter dated March 12, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to cease use, secure and quarantine affect product. Additionally, a Recall Acknowledgement Form was attached for customers to complete and return to the firm via fax to 1-800-505-4982. Contact the firm at 484-713-2197 for questions regarding this recall.
Initiated
2012-02-28
Root cause
Employee error
Terminated
2013-04-11
Distribution
Worldwide Distribution.
Recall number
Z-1622-2012
Recall status
Terminated
CDRH record ID
109173
FDA classified
2012-05-17
Recall event number
61819