Z-1622-2025: The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7

Recall record Z-1622-2025 for Maquet Cardiovascular, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1622-2025
Company in FDA record
Maquet Cardiovascular, LLC
FDA classification date
Apr 18, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.

Source facts

Action
An URGENT MEDICAL DEVICE NOTIFICATION dated March 2025 was mailed to consignees. The notification provides instructions to users on how to proceed with a procedure if a ceramic C-Ring break occurs. If a consignee has a C-Ring break they are asked to save the device and complete the provided return form so Maquet Cardiovascular can conduct an investigation. Maquet instructs consignees to forward the notification to device users and to facilities to which product was further distributed for their awareness. Consignees are asked to return the provided response form by email to recallresponses.qrc@getinge.com. Questions can be directed to Maquet/Getinge Customer Support at 888-880-2874.
Initiated
2025-03-20
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Mexico, United Arab Emirates, Singapore, India, Australia, Canada, Brazil, Hong Kong, Japan, South Korea, New Zealand.
Recall number
Z-1622-2025
Recall status
Open, Classified
CDRH record ID
212876
FDA classified
2025-04-18
Recall event number
96439

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗