Z-1624-2016: Hamilton G5 with software version between v2.41, v2.42, v2.50. Catalog number: 1590001 and 1590002. Anes

Recall record Z-1624-2016 for Hamilton Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1624-2016
Company in FDA record
Hamilton Medical, Inc.
FDA classification date
May 9, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

After performing the suctioning maneuver, including disconnecting the patient, suctioning , and reconnecting the patient, the preset pattern of ventilation many not continue as expected.

Source facts

Action
Hamilton Medical Inc. sent an" Medical Device Field Safety Corrective Action" letter dated March 7, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Hamilton Medical then had to compile a list of to whom they had distributed the affected ventilators and sent letters on April 6, 2016. Letters included the Medical Device Field Safety Corrective Action from Hamilton AG. Letters included immediate actions to be take by use and that reps will be visiting each site to upgrade ventilators to v 2.60 to alleviate the issue. Distributors will update software and will provide an updated Operator's manual as soon as it is available. Questions may be directed to 800-426-6331, Ext, 215.
Initiated
2016-03-09
Root cause
Software design
Terminated
2017-06-29
Distribution
US distribution only.
Recall number
Z-1624-2016
Recall status
Terminated
CDRH record ID
145040
FDA classified
2016-05-09
Recall event number
73816

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗