Open original FDA record ↗Recall record
- FDA identifier
- Z-1626-2021
- Company in FDA record
- Philips Respironics, Inc.
- FDA record creation date
- May 21, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.
Source facts
- Action
- In April, 2021, Philips issued an Urgent Medical Device Removal notice to customers via letter notifying them that Philips determined that certain devices were built with motor assemblies that could contain non-conforming plastic material. ACTION TO BE TAKEN BY CUSTOMER / USER 1. Locate all impacted devices listed above immediately STOP USE. All impacted devices must be removed from service. Replace with a non-impacted device if it is being used by a patient. 2. This notice needs to be passed on to all those who must be aware within your organization or to any organization where the impacted devices have been transferred. 3. Once impacted devices have been removed from service and isolated notify your Philips Representative that the impacted devices are ready for return. Your Philips Representative will arrange for return of the device. Replacements will be provided free of charge. 4. Fill out the Business Reply Form that has been provided with this letter and return it to your Philips Representative. This form serves as official acknowledgement that you have fully performed your obligations to complete this medical device recall.
- Initiated
- 2021-04-06
- Root cause
- Nonconforming Material/Component
- Terminated
- 2024-01-17
- Distribution
- Distribution US nationwide to states of: LA, PA, WI, CO, WV, WA, CA, TX, MO, OK, FL, and TN. Brazil, China, and Japan
- Recall number
- Z-1626-2021
- Recall status
- Terminated
- CDRH record ID
- 186935
- FDA record created
- 2021-05-21
- Recall event number
- 87753