Z-1626-2021: BiPAP A40, China: Continuous ventilator, non-life-supporting. Material Number - CN1111169

Recall record Z-1626-2021 for Philips Respironics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1626-2021
Company in FDA record
Philips Respironics, Inc.
FDA record creation date
May 21, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.

Source facts

Action
In April, 2021, Philips issued an Urgent Medical Device Removal notice to customers via letter notifying them that Philips determined that certain devices were built with motor assemblies that could contain non-conforming plastic material. ACTION TO BE TAKEN BY CUSTOMER / USER 1. Locate all impacted devices listed above immediately STOP USE. All impacted devices must be removed from service. Replace with a non-impacted device if it is being used by a patient. 2. This notice needs to be passed on to all those who must be aware within your organization or to any organization where the impacted devices have been transferred. 3. Once impacted devices have been removed from service and isolated notify your Philips Representative that the impacted devices are ready for return. Your Philips Representative will arrange for return of the device. Replacements will be provided free of charge. 4. Fill out the Business Reply Form that has been provided with this letter and return it to your Philips Representative. This form serves as official acknowledgement that you have fully performed your obligations to complete this medical device recall.
Initiated
2021-04-06
Root cause
Nonconforming Material/Component
Terminated
2024-01-17
Distribution
Distribution US nationwide to states of: LA, PA, WI, CO, WV, WA, CA, TX, MO, OK, FL, and TN. Brazil, China, and Japan
Recall number
Z-1626-2021
Recall status
Terminated
CDRH record ID
186935
FDA record created
2021-05-21
Recall event number
87753

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗