Open original FDA record ↗Recall record
- FDA identifier
- Z-1626-2023
- Company in FDA record
- Binx Health, Inc.
- FDA classification date
- May 23, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
May not meet the declared performance characteristics for the detection of Chlamydia trachomatis (CT) within the labeled expiry period, potential to produce either false negative CT or positive CT results
Source facts
- Action
- Binx issued Urgent Medical Device Recall Letter on 29 March 2023 via email. On 31 March 2023 a hard copy of the written Recall Notification and Response Form was sent via certified mail through the United States Postal Service. Letter states reason for recall, health risk and action to take: Discontinue use of the binx io CT/NG Assay lot listed above, and discard your inventory of this lot. " At the Clinic Director s discretion results may be reviewed. RESOLUTION: " binx health has removed this binx io CT/NG Assay lot from our inventory. We will provide alternate lots for all current and future replenishment orders or requests. Please share this information with your clinic staff and retain this notification as part of your clinic Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory or clinic, please provide them a copy of this letter. Please complete and return the enclosed Response Form within 10 days so we are assured you have received this important communication. If you have any questions regarding this notice, please contact the binx Quality Assurance department by email at qateam@mybinxhealth.com.
- Initiated
- 2023-03-29
- Root cause
- Under Investigation by firm
- Distribution
- In the US state of PA
- Recall number
- Z-1626-2023
- Recall status
- Open, Classified
- CDRH record ID
- 199776
- FDA classified
- 2023-05-23
- Recall event number
- 92143