Open original FDA record ↗Recall record
- FDA identifier
- Z-1627-2012
- Company in FDA record
- Ethicon Endo-Surgery Inc
- FDA classification date
- May 18, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There is a possibility of a component breakage, which may cause the device to stay locked on tissue after firing. If the component exhibits this issue, the device does not open either with the anvil release button or with the manual firing release lever.
Source facts
- Action
- Ethicon Endo-Surgery sent an Urgent Device Recall letter dated March 19, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory immediately to determine if they have the affected product and to remove the affected product. Fill out the Business Reply Form and fax it to 1-513-337-4138 within 3 business days, even if they do not have the affected product in their inventory. Ethicon will automatically replace any affected product returned by June 29, 2012. Customers were asked to share this information with appropriate staff. If customers should need clinical or product support call 1-800-873-3636, option 6, or contact their local sales representative. For questions regarding this recall call 513-337-7000.
- Initiated
- 2012-03-02
- Root cause
- Vendor change control
- Terminated
- 2013-03-22
- Distribution
- Worldwide Distribution - USA including AL, AR, CA, CO, FL, IA, IL, KS, LA, MD, MO, MS, NC, NM, NY, OH, SC, TX & VA; and the countries of Argentina, Ecuador, Mexico, Australia, Egypt, Singapore, Belgium, Greece, Slovenia, Brazil, India, South Africa, China, Israel, United Arab Emirates, Colombia, Japan & Venezuela.
- Recall number
- Z-1627-2012
- Recall status
- Terminated
- CDRH record ID
- 108915
- FDA classified
- 2012-05-18
- Recall event number
- 61750