Skip to content

Public FDA record

Z-1627-2020: iTotal CR Knee Replacement System, Model No. TCR-121-1111-010102, (ITOTAL CR IPOLY XE IMPLANT KIT LEFT) -

Recall record Z-1627-2020 for Conformis, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1627-2020
Company in FDA record
Conformis, Inc.
FDA record creation date
Apr 1, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A limited number of tibial trays were labeled with an incorrect serial number and were delivered to the wrong customer accounts.

Source facts

Action
Conformis Customer Service called three affected customers on February 4, 2020, to ask them to verify tibial tray serial numbers, and confirmed that the three kits in question had the incorrect tibial tray. These affected customers were asked to return the affected product to the firm. Following this action, Conformis Sales Representatives were instructed to verify that all products within their control, including those already placed in hospitals, that were manufactured between September 1, 2019, and February 3, 2020, had tibial tray serial numbers that matched the patient's serial number. No additional affected product was found in the field.
Initiated
2020-02-03
Root cause
Mixed-up of materials/components
Terminated
2020-12-02
Distribution
US Nationwide distribution in the states of LA, NY, TN.
Recall number
Z-1627-2020
Recall status
Terminated
CDRH record ID
180102
FDA record created
2020-04-01
Recall event number
85108

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗