Open original FDA record ↗Recall record
- FDA identifier
- Z-1628-2020
- Company in FDA record
- Conformis, Inc.
- FDA record creation date
- Apr 1, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A limited number of tibial trays were labeled with an incorrect serial number and were delivered to the wrong customer accounts.
Source facts
- Action
- Conformis Customer Service called three affected customers on February 4, 2020, to ask them to verify tibial tray serial numbers, and confirmed that the three kits in question had the incorrect tibial tray. These affected customers were asked to return the affected product to the firm. Following this action, Conformis Sales Representatives were instructed to verify that all products within their control, including those already placed in hospitals, that were manufactured between September 1, 2019, and February 3, 2020, had tibial tray serial numbers that matched the patient's serial number. No additional affected product was found in the field.
- Initiated
- 2020-02-03
- Root cause
- Mixed-up of materials/components
- Terminated
- 2020-12-02
- Distribution
- US Nationwide distribution in the states of LA, NY, TN.
- Recall number
- Z-1628-2020
- Recall status
- Terminated
- CDRH record ID
- 180103
- FDA record created
- 2020-04-01
- Recall event number
- 85108