Z-1629-2018: ARCHITECT DHEA-S Reagent Product Usage: It is a chemiluminescent microparticle immunoassay (CMIA) for th

Recall record Z-1629-2018 for Abbott Gmbh & Co. KG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1629-2018
Company in FDA record
Abbott Gmbh & Co. KG
FDA record creation date
May 4, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Falsely elevated results may be obtained when using the ARCHITECT DHEA-S assay with samples from infants up to 60 days old.

Source facts

Action
Customers were notified via letter on about 12/14/2017. Instructions included to immediately discontinue use of the ARCHITECH DHEA-S assay with samples form infants up to 60 days old, to immediately discontinue use of the ARCHITECH DHEA-S reagent insert Expected Values representative data for children up to 10 years old, to review the letter with the Medial Director and follow the laboratory protocol regarding the need for reviewing previously reported patient results, to forward the Product Correction letter to customers if the product was further distributed, and to complete and return the Customer Reply Form. Questions from U.S. customers can be directed to Customer Service at 1-877-4ABBOTT (available 24 hours per day, 7 days a week) and customers outside the U.S. should contact their local area Customer Service.
Initiated
2017-12-14
Root cause
Device Design
Terminated
2018-10-10
Distribution
Worldwide Distribution in the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MN, MO, MS, NC, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, VT, WA, WY. Distributed internationally to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Barbados, Belarus, Belgium, Bosnia and Herzegovi, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Egypt, Finland, France, Gaza and Jericho, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mexico, Morocco, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Yemen, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Tajikistan, Thailand, Tunisia, Turkey, Uganda, United Arab Emirates, Uganda, United Kingdom, Uruguay .
Recall number
Z-1629-2018
Recall status
Terminated
CDRH record ID
162437
FDA record created
2018-05-04
Recall event number
79482

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗