When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system
Source facts
Action
The Company will immediately notify by fax or direct contact (when available) and/or express courier each consignee purchaser and/or known user site with the laser safety notice which advises of the issue, the risk involved and how to minimize the risk of occurrence.
Initiated
2019-10-02
Root cause
Radiation Control for Health and Safety Act
Terminated
2024-07-25
Distribution
Worldwide Distribution - US Nationwide
Recall number
Z-1629-2020
Recall status
Terminated
CDRH record ID
177441
FDA record created
2020-04-06
Recall event number
84163
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.