Skip to content

Public FDA record

Z-1629-2026: Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System

Recall record Z-1629-2026 for Orthalign, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1629-2026
Company in FDA record
Orthalign, Inc
FDA classification date
Mar 24, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)

Source facts

Action
On February 10, 2026, OrthAlign issued a Urgent: Medical device Recall Notification via overnight mail. OrthAlign asked consignees to take the following actions: Please review your inventory for the affected products listed in Table 1. 1. Immediately remove and quarantine the affected products until the products are returned to OrthAlign. 2. Please share this recall notification with all members in your network as well as those where the affected devices have been transferred. 3. If you have further distributed the affected products, please identify and notify consignees. You may provide consignees a copy of this notification letter. 4. Please complete the attached response form within Five (5) business days and email the completed form to : oaregulatory@orthalign.com 5. Adverse actions are quality issues and can be reported to the Food and Drug Administration s MedWatch program: o Online: By completing and submitting the report online at www.fda.gov/medwatch o Regular mail or fax): download the form from www.fda.gov/MedWatch/getforms.htm or call 800-332-1088 to request a reporting form; then complete the form and mail it to the address on the pre-addressed form or submit by fax to 800-332-0178.
Initiated
2026-02-10
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the state of UT.
Recall number
Z-1629-2026
Recall status
Open, Classified
CDRH record ID
218966
FDA classified
2026-03-24
Recall event number
98516

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗