The pad printing around the suture entry holes on the 15mm Suture Guide included in the kit may not have been appropriately manufactured resulting in the pad
printing flaking off around the entry holes.
Source facts
Action
Consignees were notified of the recall by letter sent via FedEx dated March 20, 2015. The letter instructs consignees to do the following: discontinue use of the affected product, complete the attached acknowledgement and receipt form to return the product and schedule a replacement.
Initiated
2015-03-20
Root cause
Component design/selection
Terminated
2015-07-29
Distribution
US Nationwide distribution.
Recall number
Z-1630-2015
Recall status
Terminated
CDRH record ID
134970
FDA classified
2015-05-19
Recall event number
70828