The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.
Source facts
Action
Natus initiated the recall by letter on 03/05/2020. the firm asked customers to return the affected units to Natus Medical Incorporated, 5900 First Avenue, Seattle, WA 98108, USA. Testing will be performed by Natus on the returned unit and will be returned to the consignee following completion of testing. Natus will provide a temporary replacement unit during this time.
Initiated
2020-03-05
Root cause
Nonconforming Material/Component
Terminated
2021-08-09
Distribution
Worldwide distribution. US nationwide, Australia, Dominican Republic, United Arab Emirates, Israel, Italy, Hong Kong, Korea, Hungary, Switzerland, Colombia, and United Kingdom
Recall number
Z-1630-2020
Recall status
Terminated
CDRH record ID
180248
FDA classified
2020-04-01
Recall event number
85189